labindia logo

Tablet Dissolution Tester DS 14000+ with Piston Pump

The tablet dissolution tester DS 14000 with piston pump is a perfect solution for dissolution testing for high throughput.  Automated sampling as per pharmacopeia allows sampling tubes to be lowered in the media only at the time of sampling and withdrawn immediately thus no part of the assembly contributes motion, agitation or vibration.

This dissolution tester comes with a high-pressure piston pump that can have high flow rates up to 25ml/min with accuracy upto 1%. Different sample collection options available and all the trays have sensor to locate proper position of tray with alarm.

  There is also a recovery test facility in dissolution tester 14000 to check 100% drug dissolution. With many more features, the tablet dissolution system from labindia is one of the best in terms of price, quality, and performance.

Get Quote
Get Quote

X
Request Demo
Request Demo
X
Call Now
Tablet Dissolution Tester DS 14000 with Piston Pump Manufacturers India
Standard Features of Dissolution Tester DS 14000+ with Piston Pump
  • User-friendly, complies with current USP , IP ,BP, & EP and ASTM specifications.
  • Alphanumeric polyester soft keys for keyboard.
  • External vibration free Water Circulator for uniform water circulation with audible and low water level alarm.
  • Mono shaft design with automatic adjustment for 25 mm depth setting with easy changeover between Apparatus I & II eliminates routine height validation as per USP.
  • Paddles, Baskets and Vessels are laser marked with serial numbers for traceability.
  • Tablet dispenser drops 12 dosage form at single instance
  • Facility to monitor Vessel temp, with DTS Technology (Digital Temperature Sensor)
  • Easy placement and locking of vessels, the Ease-align system allows the vessels to simply slide into the place (Bionet Locking)
Intelligent Sampling System
  • Automated sampling as per USP Specifications. Sampling tubes are lowered in the media only at the time of sampling and withdrawn immediately after sampling, thus no part of the assembly contributes motion, agitation or vibration.
  • Sampling tubes are accurately moved to the USP sampling position i.e. a zone mid way between the sur face of media and the top of paddle/basket parameters, not less than 1 cm from the vessels wall as selected in the method.
  • 12 vessels temperature monitoring system automatically measures & records the temperature of individual vessel at specified sample points.
Pump System - PP 12
  • Piston Pump - PP 12 is ideal for sample withdrawal from 12 channels.
  • The inert ceramic rotor overcomes all absorption issue
  • Accuracy better than ±2%.(@10mL)
  • Rinse function for reducing carryover issues.
  • Surfactant Media Compatibility.
Sample Collection
  • 12 X 6 X2 sets of samples can be collected. For more sampling interval, 24 X 6 collection trays are available.
  • Option of 1.5ml & 2ml HPLC vials is available.
  • Over Head Design for electronic safety and fail safe operation.
  • Sensor to locate proper position of tray with alar m facility for collection of sample
  • Wide mouth vial to minimise SLS spillover problem due to foaming characteristics
Regulatory Compliance
  • DS 14000+ with Piston Pump meets all requirements relating to validation, qualification and calibration.
  • Appropriate qualification documents (I.Q. / O.Q.) can be supplied with the instrument.
Additional Features
  • Facility to RINSE the entire sampling path in between sampling time - point to eliminate contamination & carryover
  • Specially developed cleaning system to clean the entire sampling path after each run
  • Facilities to perform the dissolution test using two buffers (Buffer changing) to cater the application of enteric coating tablets.
  • Recovery Test facility to study 100% Drug Dissolution.
  • Split & on-time interval
Typical Specification
Parameters
DISSO SMART 14000 with Piston Pump
DS14000 with Piston Pump
Dissolution Vessel

Option for Polycarbonate / Glass Vessels (clear, amber, peak vessels, 250, 150 & 100 ml dissolution vessels available). 

Display

7" Colour High Resolution Touch Screen Display

40*2 LCD Display (Alphanumeric)

RPM

20 to 350 RPM

20 to 250 RPM

RPM Accuracy

± 1 RPM

Temprature Range

20° to 55° degree

Temp Accuracy

± 0.1° C

Method

1000

15

User 

250

10

Role

4 Role(Admin,Manager,Supervisor,Operator)

2 Role(Admin & User)

Connectivity

USB,LAN,WiFi,Bluetooth

RS-232

Test Report

10000

Last Report

21 CFR Compliance

Yes (Optional)

Yes( EDC Software)

Auditrail

Yes

NA

E Signature

Yes

NA

Sampling Interval

50

30

Sample Interval Time

1 Min to 1200 hrs

1min to 720 hrs

Pump Used for Sample Collection

Piston Pump

Rinsing

Rinse

Cleaning Cycle

Programmable multiple cleaning & drying cycles

Sample Collector Type

Covered (no evaporation, protects samples from
dust, dirt & contamination)

Open

HPLC Vials Tray

Yes

Auto Tablet Dispenser

Yes

Manual

Tray detection & identification

Special sensor for detecting the alignment and
type of tray to prevent any human error

NA

Contact parts

All Contacts Parts are inert

Sample Collection

12*8*2 Sample Collection Tray Tray.(24*8 Optional)

12*6*2 Sample Collection Tray Tray.(24*6 Optional)

Automatic Vessel Centering

Yes (Complies as per US FDA division of
pharmaceutical analysis (Ref. Document no.
DPA-LOP.002)

Center Shaft

± 0.5mm

Lift Mechanism

Motorized (Automatic) for accurate height

Heater

Isolated temperature controller (vibration-free & precise) with wake up & sleep modes (saves time & energy) (1KW)

Shaft

MonoShaft - Easy Switch Over between Paddle & Basket

Height Adjustment

Automatic height adjustment

Apparatus

Unique design of USP apparatus
I, II, V, VI & intrinsic dissolution can be used. Also Small Volume Dissolution kit 250ml & 150ml can be used.

Laser Marking on Accessories

Yes

Water Bath

Clear acrylic molded water Bath 

Water Level Sensor

Yes

Power Failure

Yes


LABINDIA reserve the right to change specification without notice as part of its continuous product development.

Software
  • Alphanumeric entries of Sample Name, Sample Number and Identification Number for authentication.
  • Built-in Real Time Clock (RTC) for date and time on display and on printout.
  • Daily Auto Incremented Run Number and factory entered CUSTOMER NAME with Instrument Serial Number on report printouts
  • Non-Volatile memory storage of 15 methods with parameters.
  • Two tier password protection of Admin and User
  • Reports can be obtained even after Resetting / Power off / Power failure conditions.
  • Error indication helps user to trace the problem.
  • Audible indication for ready state of instrument.
  • Validation Software to validate RPM, Temperature, sampling volume & replenish volume.
Tablet Dissolution Testers