UV-990 Spectrophotometer

Labindia’s UV-990 Spectrophotometer incorporates dual monochromator technology making it well suited to even the most demanding of applications in all areas of UV-Visible Spectroscopy including:
• Pharmaceutical • Material Science • Metrological Verification • Biotechnology • Food Safety – Research
The double monochromator spectrophotometer delivers extremely wide linear dynamic range by ensuring lowest stray light in contrast to a single monochromator system. This type of Instrument is suitable for measurements of highest concentration sample solutions and weak - transmittance materials, such as optical films, coated thin film, semi-conductors, glass, cellulose nanofibre, enzymatic reactions.

Get Quote
Get Quote
X
Request Demo
Request Demo
X
Call Now
Call On: 9136909578
Sales Support: sales.mfd@labindia.com
Labindia’s UV-990 Spectrophotometer incorporates dual monochromator technology making it well suited
Standard Features of UV-990 Spectrophotometer
  • Truly double beam Double Monochromatic Design with holographic grating 1800 lines/mm
  • Variable Bandwidth with 0.1 to 5 nm.
  • Offer least Signal to noise ratio- 0.00005Abs RMS (500nm?
  • Integrating sphere with 60/100 mm diameter
  • Ultra Low Stray Light  <   0.0003%T (Nal at 220 nm, NaNO2 at 340 nm) Higher Accuracy, Broader Linear Dynamic Range.
Typical Specification

<!--[if gte mso 9]><xml> </xml><![endif]--><!--[if gte mso 9]><xml> Normal 0 false false false EN-IN X-NONE X-NONE </xml><![endif]--><!--[if gte mso 9]><xml> </xml><![endif]--><!--[if gte mso 10]> <style> /* Style Definitions */ table.MsoNormalTable {mso-style-name:"Table Normal"; mso-tstyle-rowband-size:0; mso-tstyle-colband-size:0; mso-style-noshow:yes; mso-style-priority:99; mso-style-parent:""; mso-padding-alt:0cm 5.4pt 0cm 5.4pt; mso-para-margin-top:0cm; mso-para-margin-right:0cm; mso-para-margin-bottom:8.0pt; mso-para-margin-left:0cm; line-height:107%; mso-pagination:widow-orphan; font-size:11.0pt; font-family:"Calibri","sans-serif"; mso-ascii-font-family:Calibri; mso-ascii-theme-font:minor-latin; mso-hansi-font-family:Calibri; mso-hansi-theme-font:minor-latin; mso-fareast-language:EN-US;} table.MsoTableGrid {mso-style-name:"Table Grid"; mso-tstyle-rowband-size:0; mso-tstyle-colband-size:0; mso-style-priority:39; mso-style-unhide:no; border:solid windowtext 1.0pt; mso-border-alt:solid windowtext .5pt; mso-padding-alt:0cm 5.4pt 0cm 5.4pt; mso-border-insideh:.5pt solid windowtext; mso-border-insidev:.5pt solid windowtext; mso-para-margin:0cm; mso-para-margin-bottom:.0001pt; mso-pagination:widow-orphan; font-size:11.0pt; font-family:"Calibri","sans-serif"; mso-ascii-font-family:Calibri; mso-ascii-theme-font:minor-latin; mso-hansi-font-family:Calibri; mso-hansi-theme-font:minor-latin; mso-fareast-language:EN-US;} </style> <![endif]-->

Model

Specifications

Optical system

Double beam, Double Monochromator, Czerny Turner

Bandwidth

0.1/0.2/0.5/1.0/2.0/4.0/5.0 nm

Wavelength range

190-900nm

Wavelength accuracy

±0.1nm (D2656.1nm),±0.3nm all range

Wavelength repeatability

≤0.02nm

Wavelength display

0.1nm

Resolution

0.1nm

Wavelength scan rate

4000 to 0.5 nm/min

Wavelength slew rate

14500 nm/min

Photometric accuracy

±0.002 A @ 0.0-0.5 A ,+ 0.004 A @ 0.5-1 A, ±0.3%T @ 0-100 %T

Photometric repeatability

< 0>

Baseline flatness

±0.0008 A

Baseline stability

±0.0004Abs (200-860nm)

Noise level

< 0>

Photometric range

- 5.0 to +5.53 Abs

Stray light

< 0>

Grating

Holographic Gratings 1800 lines/mm

Detector

PMT

Light source

D2/WI

Optional Accessory

Specular Reflection Accessory, Integrating Sphere 60mm / 100mm Automatic 8 Cell Changer, long path cell holder, Peltier Solid Sample holder etc.

Interface

LAN, USB, RS-232

Power supply

160W

Dimensions

660 (W) x 570 (D) x 270 (H) mm

Weight

32 kg