UV-990 Spectrophotometer

Labindia’s UV-990 Spectrophotometer incorporates dual monochromator technology making it well suited to even the most demanding of applications in all areas of UV-Visible Spectroscopy including:
• Pharmaceutical • Material Science • Metrological Verification • Biotechnology • Food Safety – Research
The double monochromator spectrophotometer delivers extremely wide linear dynamic range by ensuring lowest stray light in contrast to a single monochromator system. This type of Instrument is suitable for measurements of highest concentration sample solutions and weak - transmittance materials, such as optical films, coated thin film, semi-conductors, glass, cellulose nanofibre, enzymatic reactions.

Get Quote
Get Quote
X
Request Demo
Request Demo
X
Call On: 9136909578
Sales Support: sales.mfd@labindia.com
Labindia’s UV-990 Spectrophotometer incorporates dual monochromator technology making it well suited
Standard Features of UV-990 Spectrophotometer
  • Truly double beam Double Monochromatic Design with holographic grating 1800 lines/mm
  • Variable Bandwidth with 0.1 to 5 nm.
  • Offer least Signal to noise ratio- 0.00005Abs RMS (500nm?
  • Integrating sphere with 60/100 mm diameter
  • Ultra Low Stray Light  <   0.0003%T (Nal at 220 nm, NaNO2 at 340 nm) Higher Accuracy, Broader Linear Dynamic Range.
Typical Specification
Parameters
UV-990(UV-Visible)
Photometric Range

- 6.0 to + 6.0 Abs

Stray Light

  <   0.0003%T (NaI at 220 nm, NaNO2 at 340 nm)

Optical System

Double beam, double monochromator

Wavelength  Range

190-900nm

Wavelength Accuracy

±0.1nm (D2656.1nm),± 0.3nm all range
 

Wavelength Reproducibility

  <  0.05nm

Spectral Bandwidth

0.1/0.2/0.5/1.0/2.0/4.0/5.0 nm

Detector

PMT

Light Source

D2/WI

User Interface

 

Baseline Flatness

±0.0008 A
 

Scan speed

 

Cell Capacity

 

Interface

LAN, USB, RS-232

Wavelength display

0.1nm

Grating Holographic Gratings

1800 lines/mm

Power Supply

160W

Dimesions

660 (W) x 570 (D) x 270 (H) mm

Weight

32kg

Baseline stability

±0.0004Abs (200-860nm)
 

Photometric accuracy

±0.002 A   @ 0.0-0.5 A,  + 0.004 A @ 0.5-1 A ±0.3%T @ 0-100 %T
 

Photometric repeatability

  <  0.001 A @ 0.0-0.5 A,  <  0.002 A @ 0.5-1 A,  <  0.15 %T @ 0-100 %T

Noise level

  <  0.00005 A / h @ 500nm, 2 hours

 

LABINDIA reserve the right to change specification without notice as part of its continuous product development.